DOM MB Diagnosis page

Board ID:

Rev 5.0 SN 000340

Source:

LED Testing; integration step 5

Symptoms:                            

Light pulse count lower than expected

Analyst(s):

G. Przybylski

Analysis Procedure(s):

U6 output at CR1 doesn't change when DAC 12 is changed.  Suspect U6 defective.

Root cause:

Defective U6, the switching controller chip.

Mitigation:

Replace defective part. 

Disposition:

Mainstream

Any Other Information:

Traveler