Board ID:
Rev 5.0 SN 000537
Source:
LED Testing; integration step
Symptoms:
Light pulse count lower than
expected
Analyst(s):
G. Przybylski
Analysis Procedure(s):
Tested LED power circuit: ledon/ledmax/ledmin. Voltages normal. LED
Waveforms normal. DAC control of amplitude works properly. Low
count suspected in event analysis algorithm.
Root cause:
Unclear. Possible testing issue
Mitigation:
Unclear. Possible testing
issue.
Disposition:
Mainstream
Any Other Information:
Traveler