DOM MB V5 Diagnosis
page
Board ID:
V5.0L 000626
Source:
Azriel Goldschmidt
Symptoms:
September 10, Boot failure cold, and ATWD A Channel 2 Baseline
Jumps when cold
Analyst(s):
G. Przybylski
Analysis Procedure(s):
Ran low temperature profile from -40°C
to -75°C. Results:
Local Coincidence Loopback failure at and below -65°C.
ATWD A Ch 1 Baseline failure only at -50°C, and at -55°C.
Interesting that the problem doesn't persist to -75°C.
Root cause:
Boot race
Cause of Baseline problem is uncertain
Mitigation:
After firmware upgrade, Pass by fiat.
Disposition:
Mainstram with restriction that the board be integrated in Europe
to insure it cannot become an IceTop module.
Any Other Information:
Traveler September 10