DOM MB V5.1 Diagnosis page

Board ID:

V5.1 001344

Step:

Vendor STF Test

Symptoms:

April 21, 2005,  Pressure reading incorrect

Analyst(s):

G. Przybylski

Analysis Procedure(s):

5/4/05: Probed U1-4 while grounding U1-8.  Reading is 107mV instead of the expected ~4.0V.

Root cause:

Faulty U1

Mitigation:

Replace, retest and mainstream

Disposition:

Repair pending 3/4/05

Any Other Information:

Traveler April 21, 2005