DOM MB V5.1 Diagnosis page

Board ID:

V5.1 001446

Step:

Vendor: Room Temperature STF test

Symptoms:

April 19, 2005, ATWD Pulser failures for All ATWD tests
See failures

Analyst(s):

G. Przybylski

Analysis Procedure(s):

5/6/05: Confirmed STF test failures.  Probed pulser circuit. Trigger OK, output OK. Probed delay line. Found incorrect readings.

Root cause:

Faulty delay line.

Mitigation:

Replaced delay line,  Passes rom temperature STF Tests.

Disposition:

Mainstream

Any Other Information:

Traveler April 19, 2005